Lab World / instrument integration / Di Hu
9 stations. For each one: is it already decoded, how does software reach it, and what does a first run look like.
Every run on this page is a simulation. Nothing here ran on a physical instrument. Built against the public PyLabRobot source at commit 182f92e of 23 Nov 2026.
TODO_1, TODO_2 and so on are placeholders used on purpose: each marks a payload not yet decoded or a fact about the real instrument not yet known.
Steps 3 to 5 follow the PyLabRobot device driver guide. Steps 3 and 6 follow my reverse engineering playbook.
| Station | Where it stands | How software reaches it | In simulation |
|---|---|---|---|
| Tissue-culture station | Instrument not chosen yet | Not decided. The example uses a made-up stand-in: 1 JSON message per release, the adapter speaks first, the station acknowledges or refuses. A real adapter would wrap what the chosen supplier documents. | 5 passed |
| Agilent BioTek EL406 | Decoded, in PyLabRobot | FTDI USB, 38400 baud 8N2, no flow control. Host speaks first: 11-byte header (cmd, length, checksum), payload (protocol.py:15-25). Reply: ACK 0x06, header, data (el406.py:318-319). Steps polled by 0x92 (el406.py:595). | 4 passed |
| Hamilton Microlab STAR | Decoded, in PyLabRobot | USB, VID 0x08AF, PID 0x8000 (master.py:81-82). The host speaks first: ASCII commands like C0TPid0117xp... Each reply repeats module, command and id, then er00/00 or an error code (framing.py:3-5 and framing.py:319-320). | 6 passed |
| LiCONiC STX44 incubator | Decoded, in PyLabRobot, older interface | RS-232, 9600 baud 8E1, RTS/CTS (liconic_backend.py:80-90). Host speaks first: a break, CR answered by CC, then 1 ASCII command per line (ST, RS, RD, WR DMn) ending in CR; replies end CR LF (liconic_backend.py:96-139). | 4 passed |
| Rail-mounted transfer arm (linear track) | Maker's SDK, nothing to decode | Through the xArm controller with xarm-python-sdk: the same XArmAPI object as the arm calls get/set_linear_motor_*; the SDK sends them as Modbus to the control box RS-485 (xarm-python-sdk xarm/x3/linear_motor.py:40-46). | 6 passed |
| UFACTORY xArm 6 plate arm | Maker's SDK, nothing to decode | Ethernet to the xArm controller through xarm-python-sdk (hello-world.ipynb:12). XArmAPI(ip) is created in setup(), then each operation is 1 blocking SDK call run in a thread (xarm6.py:86). | 7 passed |
| Passive plate hotel | Nothing to decode | Nothing to reach: no wire. Software keeps a model of what sits on each shelf (PyLabRobot resources) and the arm that loads it is the only actor. The scheduler owns reservations and timers. | 5 passed |
| BMG LABTECH PHERAstar FSX | Not decoded yet | Unknown. Upstream's 2 BMG readers share 1 framing module: the CLARIOstar speaks FTDI frames (STX, size, header, payload, 24-bit sum, CR) or USB packets, the VANTAstar USB only. The host asks, then polls status. | 7 passed |
| Shared LC-MS bench | Nothing to decode | Rear remote connector: software reads a Ready line and closes a Start contact through opto-isolators (README.md:88-97). The vendor software holds the method. Raw files are picked up from a data folder. | 5 passed |
Station 1 of 9 / Instrument not chosen yet
Supplier and interface are still open, so there is no protocol to show yet.
Release 1 cell batch for seeding and check that the receiving worklist names that batch in every row. Batch and worklist ids are made up.
adapter = CultureAdapter(station)
worklist = [{"plate": "seed_plate_1", "well": w, "batch": "CB-0001"}
for w in ("A1", "A2", "A3")]
log("ask release CB-0001 for seeding against worklist WL-7")
try:
adapter.release("CB-0001", "WL-7", worklist)
except ReleaseNotConfirmed as e:
out.update(outcome="hold", evidence=str(e))
except ReleaseRefused as e:
out.update(outcome="error", evidence=str(e))
else:
out.update(outcome="completed", evidence="CB-0001 released, worklist WL-7 matches")ask release CB-0001 for seeding against worklist WL-7
send {"op": "release", "batch": "CB-0001", "worklist": "WL-7"}
recv {"ok": true, "batch": "CB-0001"}
state released {'CB-0001': 'WL-7'}, executed 1
done completed: CB-0001 released, worklist WL-7 matchesSources / PREFLIGHT.md / APPROACH.md
Station 2 of 9 / Decoded, in PyLabRobot
In PyLabRobot, v1 class EL406, registry status mostly (devices.json:120-122). Binary frames over FTDI USB.
Seed cells into a 384-well plate: set up, open a batch, prime the peristaltic lines with 500 uL, dispense 40 uL into every well (24 columns), close the batch. Made-up round volumes.
el406 = EL406(name="el406")
el406.io = sim # the only swap: the FTDI transport becomes the simulator
tap.ask("set up: FTDI 38400 8N2, TEST_COMM, INIT_STATE, RESET")
await el406.setup()
plate = cellvis_384_wellplate_120uL_Fb(name="cells_384")
el406.set_plate(plate)
tap.ask("open a batch for the 384-well plate")
async with el406.batch():
tap.ask("prime the peristaltic lines, 500 uL")
await el406.peristaltic_dispenser.prime(volume=500)
tap.ask("dispense 40 uL per well, columns 1-24")
volumes = {column: 40 for column in range(1, 25)} # uL per well, by column
await el406.peristaltic_dispenser.dispense(volumes=volumes)
tap.ask("close the batch: home motors, end-of-batch marker")
await el406.stop()ask set up: FTDI 38400 8N2, TEST_COMM, INIT_STATE, RESET
send 3 frames, by command: TEST_COMM INIT_STATE RESET
recv 3 replies, last: 06 01 02 70 00 01 00 00 00 00 8c ff
state plate -, batch closed, prime 0 uL, 0 cols at [] uL, err 0x0000
ask open a batch for the 384-well plate
send 7 frames, by command: 0x00bf 0x00c1 0x00f2 0x00f4 0x0154 0x0102 0x010a
send 01 02 8d 00 01 00 00 01 00 6d ff 01
recv 8 replies, last: 06 01 02 8d 00 01 00 00 00 00 6f ff
state plate 384 std, batch open, prime 0 uL, 0 cols at [] uL, err 0x0000
ask prime the peristaltic lines, 500 uL
send 1 frames, by command: STATUS_POLL
send 01 02 90 00 01 00 00 0b 00 67 fe 01 f4 01 00 00 02 01 00 01 00 00
send 2 frames, by command: STATUS_POLL STATUS_POLL
recv 4 replies, last: 06 01 02 92 00 01 00 00 09 00 60 ff 00 00 01 00 00 00 00 00 00
state plate 384 std, batch open, prime 500 uL, 0 cols at [] uL, err 0x0000
ask dispense 40 uL per well, columns 1-24
send 1 frames, by command: STATUS_POLL
send 01 02 8f 00 01 00 00 18 00 d2 fb 01 28 00 02 00 00 00 4d 01 0a 00 02 ff ff ff 00 00...
send 2 frames, by command: STATUS_POLL STATUS_POLL
recv 4 replies, last: 06 01 02 92 00 01 00 00 09 00 60 ff 00 00 01 00 00 00 00 00 00
state plate 384 std, batch open, prime 500 uL, 24 cols at [40] uL, err 0x0000
ask close the batch: home motors, end-of-batch marker
send 2 frames, by command: HOME_MOTORS END_OF_BATCH
recv 2 replies, last: 06 01 02 8c 00 01 00 00 00 00 70 ff
state plate 384 std, batch closed, prime 500 uL, 24 cols at [40] uL, err 0x0000
done completed: sim: 24/24 columns at 40 uL/well, peristaltic pump, plate byte 1 (384 std)Sources / el406.py at 182f92e / protocol.py (framing, checksum) / peristaltic_dispenser.py / devices.json registry entry / Legacy wash tests (reference captures)
Station 3 of 9 / Decoded, in PyLabRobot
In PyLabRobot. The real STARDriver sends text firmware commands over USB.
Start-up conversation, then pick up 8 tips, aspirate 50 uL from column 1 of a source plate, dispense into column 1 of a destination plate, return the tips. Volumes made up.
def build(link):
deck = STARLetDeck()
tip_car = hamilton_tip_carrier_L5(name="tip_carrier")
tip_car[0] = tips = hamilton_96_tiprack_300uL(name="tips")
deck.assign_child_resource(tip_car, track=1)
plt_car = hamilton_plate_carrier_L5_ac(name="plate_carrier")
plt_car[0] = source = cor_96_wellplate_360uL_Fb(name="source")
plt_car[1] = dest = cor_96_wellplate_360uL_Fb(name="destination")
deck.assign_child_resource(plt_car, track=8)
# On hardware: STARDriver() with no io opens the USB device itself.
star = STARlet(deck=deck, driver=STARDriver(io=link))
return star, tips, source, dest
async def sample_prep(star, tips, source, dest):
await star.setup()
await star.pipettes.pick_up_tips(tips["A1:H1"])
await star.pipettes.aspirate(source["A1:H1"], [50.0] * 8)
await star.pipettes.dispense(dest["A1:H1"], [50.0] * 8)
await star.pipettes.return_tips()ask set up; 8 tips; 50 uL from source col 1 to destination col 1; return tips
send C0RMid0001
recv C0RMid0001er00/00kb01kp08
send C0VIid0034
recv C0VIid0034er00/00
send C0DIid0036xp08000&yp4050 3783 3516 3249 2982 2715 2448 2181tp2450tz1220te3600tm1 1 1 1 1 1 ...
recv C0DIid0036er00/00
send C0TPid0117xp01179 01179 01179 01179 01179 01179 01179 01179yp1458 1368 1278 1188 1098 1008 ...
state tips on 8 of 8 channels
recv C0TPid0117er00/00
send C0ASid0176at0 0 0 0 0 0 0 0tm1 1 1 1 1 1 1 1xp02758 02758 02758 02758 02758 02758 02758 027...
state left in source wells: [44.9] uL each; tips hold [55.1] uL each
recv C0ASid0176er00/00
send C0DSid0227dm2 2 2 2 2 2 2 2tm1 1 1 1 1 1 1 1xp02758 02758 02758 02758 02758 02758 02758 027...
state now in destination wells: [55.1] uL each; tips hold [0.0] uL each
recv C0DSid0227er00/00
send C0TRid0238xp01179 01179 01179 01179 01179 01179 01179 01179yp1458 1368 1278 1188 1098 1008 ...
state tips on 0 of 8 channels, 96 in the rack
recv C0TRid0238er00/00
done completed: destination col 1: [55.1] uL each (50 uL asked + liquid class); 96 tips in rackSources / Installation and USB interface / Debugging STAR issues (VENUS trace files) / STAR design notes / STAR recordings README / Device registry / STARDriver (master.py) / Reply framing / Legacy STAR tests (captured replies) / STAR liquid classes (mapping.py)
Station 4 of 9 / Decoded, in PyLabRobot, older interface
In PyLabRobot legacy only (v0); STX44_IC is a listed model (constants.py:5). ASCII over RS-232.
Take 1 plate from the transfer station into cassette 1, position 1, fetch it back out to the transfer station, then read the temperature. Made-up slot and a made-up 37.0 deg C.
backend = ExperimentalLiconicBackend(model="STX44_IC", port="/dev/ttyUSB0")
backend.io = sim # the only swap: the Serial transport becomes the simulator
rack = liconic_rack_17mm_22(name="cassette_1") # 22 positions, 23 mm pitch
stx = Incubator(backend=backend, name="stx44", size_x=600, size_y=600, size_z=700,
racks=[rack], loading_tray_location=Coordinate(0, 0, 0))
tap.ask("set up: break, CR handshake, activate handling, wait ready")
await stx.setup()
plate = cellvis_384_wellplate_120uL_Fb(name="cells_384")
stx.loading_tray.assign_child_resource(plate)
tap.ask("take the plate in from the transfer station to cassette 1, position 1")
await stx.take_in_plate(rack.sites[0])
tap.ask("fetch the plate back out to the transfer station")
await stx.fetch_plate_to_loading_tray("cells_384")
tap.ask("read the temperature")
temperature = await stx.get_temperature()ask set up: break, CR handshake, activate handling, wait ready
send CR | ST 1801 | RD 1915
recv CC | OK | 1
state plate at transfer, ready 1, error 00000, temp 37.0 deg C
ask take the plate in from the transfer station to cassette 1, position 1
send WR DM0 1 | WR DM23 788 | WR DM25 22 | WR DM5 1 | ST 1904 | RD 1915 | ST 1903
recv OK x5 | 1 | OK
state plate at cassette 1/1, ready 1, error 00000, temp 37.0 deg C
ask fetch the plate back out to the transfer station
send WR DM0 1 | WR DM23 788 | WR DM25 22 | WR DM5 1 | ST 1905 | RD 1915 | ST 1903
recv OK x5 | 1 | OK
state plate at transfer, ready 1, error 00000, temp 37.0 deg C
ask read the temperature
send RD DM982
recv 370
state plate at transfer, ready 1, error 00000, temp 37.0 deg C
state upstream warning: Liconic racks need to be configured manually on each setup
done completed: sim: plate path shovel > cassette 1/1 > shovel > transfer; 37.0 deg CSources / liconic_backend.py at 182f92e / constants.py (models, error codes) / liconic_backend_tests.py / Legacy Incubator frontend / Device driver guide / devices.json registry entry
Station 5 of 9 / Maker's SDK, nothing to decode
Not in PyLabRobot yet, the maker's SDK documents it. The example writes the class.
Home the track if it is not homed, travel to a taught station at 450 mm and 200 mm/s, confirm the arrival from the position report. Positions and speed are made up.
async def travel(track: LinearTrack) -> int:
await track.setup() # enables the track if it was off
await track.home() # no-op when already homed
at = await track.move_to(READER_STATION_MM, speed=SPEED_MM_S)
await track.stop() # restores the enable state found at setup
return at
# track = LinearTrack(ip="192.168.1.220", travel_mm=700)ask setup, home, move to 450 mm at 200 mm/s, stop
send get_linear_motor_is_enabled()
recv (0, 0)
send set_linear_motor_enable(True)
recv 0
send get_linear_motor_on_zero()
recv (0, 0)
send set_linear_motor_back_origin(wait=False,auto_enable=True)
recv 0
send get_linear_motor_on_zero()
recv (0, 1)
send get_linear_motor_pos()
recv (0, 0)
send set_linear_motor_pos(450,speed=200,wait=False)
recv 0
send get_linear_motor_error()
recv (0, 0)
send get_linear_motor_status()
recv (0, 0)
send get_linear_motor_pos()
recv (0, 450)
send set_linear_motor_enable(False)
recv 0
state pos 450 mm, enabled 0, on_zero 1, error 0, moves 1
done completed: position report 450 mm == taught 450 mm, error register 0Sources / PyLabRobot device-driver guide / xArm Python SDK API doc, master branch (linear motor calls)
Station 6 of 9 / Maker's SDK, nothing to decode
In PyLabRobot (XArm6). It wraps the maker's published Python SDK, so there is no wire to decode.
Carry 1 plate from a hotel nest to the reader nest: above pick, open, descend, close, lift, above place, descend, open, retreat. All poses are made up.
async def carry(arm: XArm6) -> None:
up = Coordinate(z=80)
down = Rotation(x=180, y=0, z=0)
await arm.move_to_location(HOTEL + up, down)
await arm.move_gripper(width=arm.max_gripper_width)
await arm.pick_up_at_location(HOTEL, down, resource_width=80)
await arm.move_to_location(HOTEL + up, down)
await arm.move_to_location(READER + up, down)
await arm.drop_at_location(READER, down, resource_width=80)
await arm.move_to_location(READER + up, down)
# arm = XArm6(ip="192.168.1.220"); await arm.setup(); await carry(arm)ask setup()
state setup made 9 SDK calls, all returned 0; mode 0, state 0, gripper enabled True
ask carry 1 plate: hotel nest -> reader nest
send set_position(x=300,y=200,z=180,roll=180,pitch=0,yaw=0,speed=100,mvacc=2000,wait=True)
recv 0
send set_gripper_position(850,wait=True,speed=0)
recv 0
send set_position(x=300,y=200,z=100,roll=180,pitch=0,yaw=0,speed=100,mvacc=2000,wait=True)
recv 0
send set_gripper_position(97,wait=True,speed=0)
recv 0
send set_position(x=300,y=200,z=180,roll=180,pitch=0,yaw=0,speed=100,mvacc=2000,wait=True)
recv 0
send set_position(x=300,y=-200,z=130,roll=180,pitch=0,yaw=0,speed=100,mvacc=2000,wait=True)
recv 0
send set_position(x=300,y=-200,z=50,roll=180,pitch=0,yaw=0,speed=100,mvacc=2000,wait=True)
recv 0
send set_gripper_position(850,wait=True,speed=0)
recv 0
send set_position(x=300,y=-200,z=130,roll=180,pitch=0,yaw=0,speed=100,mvacc=2000,wait=True)
recv 0
state plate at reader, pose [300, -200, 130], error_code 0
done completed: plate at reader nest, jaws at 850 of 850 units, error_code 0Sources / xarm6.py / xarm6_tests.py / xArm 6 hello-world notebook / devices.json
Station 7 of 9 / Nothing to decode
No controller, so nothing to decode. In PyLabRobot it is a resource: plate holders on a carrier.
Reserve shelf 5, place 1 skirted 384-well assay plate, wait a 600 s dwell on a simulated clock, pick it up, free the shelf. Sizes, pitch and dwell are made up.
hotel_sched = Scheduler(hotel, sim, clock)
plate = Greiner_384_wellplate_28ul_Fb("assay_plate_1")
log("ask reserve slot 5, place, dwell 600 s, pick up")
hotel_sched.reserve(5, plate.name)
try:
hotel_sched.place(5, plate) # hotel[5] = plate
except ValueError as e: # "spot 5 already has a resource" (carrier.py:86)
out.update(outcome="error", evidence=f"ValueError: {e}")
else:
clock.advance(DWELL_S)
log(f"state t={clock.now_s:g} s, dwell done: {hotel_sched.dwell_done(5)}")
if hotel_sched.pick_up(5):
out.update(outcome="completed", evidence="pickup confirmed, slot 5 free")
else:
out.update(outcome="hold", evidence=f"pickup sent {sim.pickups} time, no "
"confirmation; slot 5 stays reserved")ask reserve slot 5, place, dwell 600 s, pick up
send place slot=5 plate=assay_plate_1
recv place ok slot=5
state t=600 s, dwell done: True
send pickup slot=5 plate=assay_plate_1
recv pickup ok slot=5
state model slot 5: None; shelf 5: None; reserved: []
done completed: pickup confirmed, slot 5 freeSources / carrier.py / resource_holder.py / greiner/plates.py (384-well plate used)
Station 8 of 9 / Not decoded yet
Not in PyLabRobot. Upstream drives 2 other BMG readers; this one's commands are unknown.
A stand-in for the vendor software moves the tray out and in and reads a 384-well plate on a synthetic reader; the frames are captured, decoded, replayed behind guards, then read by a draft class.
reader = PHERAstarDraft(transport=transport) # unverified, commands are TODO_n
await reader.setup() # logs a loud "unverified" warning
try:
await reader.ensure_tray_in() # reads state, moves only if needed, confirms
values = await reader.read_plate() # 1 read command, never re-sent
outcome = "completed"
evidence = f"{len(values)} made-up values, well A1 = {values[0]}"
except ReplyLost as e: # read sent, no reply: stop here, a human checks
outcome, evidence = "hold", str(e)
except ReaderDeviceError as e: # error bit set in the reply
outcome, evidence = "error", f"device error codes {e.codes} (synthetic code)"
finally:
await reader.stop()ask (a) capture: vendor stand-in does 3 actions, 1 at a time, each marked
state captured tray_out: TODO_1 + TODO_4; tray out, plate on tray, plates read 0
state captured tray_in: TODO_2 + TODO_4; tray in, plate on tray, plates read 0
state captured read_plate: TODO_3 + TODO_4; tray in, plate on tray, plates read 1
ask (b) decode: framing, then group frames by action mark
state 6 host frames fit STX|size|hdr|payload|sum24|CR, header 0x3f; Modbus RTU frames found: 0
state in every window (status candidate): b'TODO_4'
state tray_out vs tray_in differ at offsets [9, 12]
state protocol map coverage 4/4
ask (c) replay TODO_3 (read) behind plr_re guards: armed, actuation allowed (sim)
send 02 00 0e 3f 54 4f 44 4f 5f 33 00 02 17 0d
recv 02 03 0c 3f 7d 03 02 00 03 e8 04 0d 04 32 04 57 04 7c 04 a1 04 c6 04 eb ... (780 bytes)
state replay: 384 values back; tray in, plate on tray, plates read 2
ask (d) the same read through the draft driver class
send 02 00 0e 3f 54 4f 44 4f 5f 34 00 02 18 0d
recv 02 00 12 3f 01 03 02 00 05 00 00 00 00 00 00 00 5e 0d
send 02 00 0e 3f 54 4f 44 4f 5f 33 00 02 17 0d
recv 02 03 0c 3f 7d 03 02 00 03 e8 04 0d 04 32 04 57 04 7c 04 a1 04 c6 04 eb ... (780 bytes)
state tray in, plate on tray, plates read 3; executed in (d): {'TODO_3': 1}
done completed: 384 made-up values, well A1 = 1000Sources / BMG LABTECH workflow automation / Upstream BMG framing / Upstream VANTAstar reply format / Upstream device-driver guide / Upstream device registry / plr-re capture and marks / plr-re decode / plr-re guarded replay / plr-re byte connections / plr-re protocol map
Station 9 of 9 / Nothing to decode
Not in PyLabRobot. The toolkit playbook's route, a Ready line and a Start line, needs no decoding.
8 aliquots from 1 plate as an injection sequence: for each, check Ready, log the dry-run Start (nothing sent), wait for its raw file. A missing file holds; Ready low stops with the fault text.
remote = Agilent6530Remote(PINS, Guards(armed=False), io=station) # dry run
for row, aliquot in enumerate(aliquots, 1):
if not remote.is_ready(): # reads the Ready line
outcome = "error"
evidence = f"Ready low before row {row}: {station.fault_text}; no Start sent"
break
remote.start_run() # dry run: logs the Start pulse, sends nothing
for note in notes.notes:
log(f"send {note}")
station.on_note(note) # the simulated workstation acts on it
notes.notes.clear()
raw = await wait_for_raw_file(raw_dir, row, aliquot)
if raw is None: # started, no file: hold, never start this aliquot again
outcome = "hold"
evidence = f"row {row} {aliquot} started, no raw file; not re-injected"
break
matched[aliquot] = raw
log(f"recv {raw}")ask run sequence.csv: 8 injections from PLATE01, 1 start each
send [dry-run] would pulse START on BCM27 for 0.25s
recv 001_PLATE01-A1.raw.stub
send [dry-run] would pulse START on BCM27 for 0.25s
recv 002_PLATE01-B1.raw.stub
send [dry-run] would pulse START on BCM27 for 0.25s
recv 003_PLATE01-C1.raw.stub
send [dry-run] would pulse START on BCM27 for 0.25s
recv 004_PLATE01-D1.raw.stub
send [dry-run] would pulse START on BCM27 for 0.25s
recv 005_PLATE01-E1.raw.stub
send [dry-run] would pulse START on BCM27 for 0.25s
recv 006_PLATE01-F1.raw.stub
send [dry-run] would pulse START on BCM27 for 0.25s
recv 007_PLATE01-G1.raw.stub
send [dry-run] would pulse START on BCM27 for 0.25s
recv 008_PLATE01-H1.raw.stub
state ready True, rows started 8/8, raw files 8; starts per aliquot max 1
done completed: 8/8 injections matched to a raw fileSources / Upstream device registry / Toolkit playbook (LC-MS example) / Toolkit playbook wiring / Toolkit contact-closure controller / Toolkit tests / Toolkit README / Toolkit example pin map
The central harness
Every station answers the harness the same way: completed, hold, or error.
The first run is clean. In the second run the reader carries out the read and its reply is lost. The harness holds, sends no second read, and does not ask the next station.
run SIM-001, clean: 18 record lines in harness/runs/SIM-001.jsonl
1 culture completed CB-0001 released, worklist WL-7 matches
2 seed completed sim: 24/24 columns at 40 uL/well, peristaltic pump, plate byte 1 (384 std)
3 star completed destination col 1: [55.1] uL each (50 uL asked + liquid class); 96 tips i...
4 inc completed sim: plate path shovel > cassette 1/1 > shovel > transfer; 37.0 deg C
5 rail completed position report 450 mm == taught 450 mm, error register 0
6 xarm completed plate at reader nest, jaws at 850 of 850 units, error_code 0
7 hotel completed pickup confirmed, slot 5 free
8 reader completed 384 made-up values, well A1 = 1000
9 lcms completed 8/8 injections matched to a raw file
run SIM-002, the reader's reply is lost: 16 record lines in harness/runs/SIM-002.jsonl
1 culture completed CB-0001 released, worklist WL-7 matches
2 seed completed sim: 24/24 columns at 40 uL/well, peristaltic pump, plate byte 1 (384 std)
3 star completed destination col 1: [55.1] uL each (50 uL asked + liquid class); 96 tips i...
4 inc completed sim: plate path shovel > cassette 1/1 > shovel > transfer; 37.0 deg C
5 rail completed position report 450 mm == taught 450 mm, error register 0
6 xarm completed plate at reader nest, jaws at 850 of 850 units, error_code 0
7 hotel completed pickup confirmed, slot 5 free
8 reader hold no reply to b'TODO_3'; not re-sent, reader state unknown
9 lcms not asked waiting behind reader19 harness tests passed: each of the 9 stations with a lost reply and with an instrument error.
Run it yourself
uv run pytest # all 79 tests
uv run python -m stations.star.drive # 1 station, with its wire traffic
uv run python -m harness.walk # 1 plate through all 9To watch the runs step by step, open
replay.html. It plays the trace files that
uv run python -m harness.trace records.
Read this before quoting the page