Agentic Lab Guide Liquid handling  ·  Di Hu

A part of its own · for any liquid handler

Liquid handling at the scientist's bar.

An expert's guide to tuning a liquid handler: the blowout volume, the speed and the distance from the bottom. Use it as a starting principle for what would work, then tune it on your own instrument.

The second pass

Go into the deck.

The first pass of a protocol gets the deck layout, the protocol and the basic steps. Then you go into the deck and think it through. When you do this experiment, with the tips in place, how do you want this to go?

By the physics of the tip, the liquid might hang as a drop. So you want a certain degree of blowout volume, a certain degree of speed, and a certain distance from the bottom. All of these are tunable parameters.

The tunable parameters

Where to start.

Start with regular parameters, maximum speed for everything. Then tune:

Tunable parameterWhat the expert wantsStart here
Blowout volumeSufficient blowout, to ensure proper liquid handling.For low volume, 20 percent of the volume.
SpeedMaximum speed for everything. Gentler into cells. A forceful dispense for lysis.Maximum. Into cells, 50 to 70 percent of maximum.
Distance from the bottom, at aspirateLow enough that the liquid is not missed.Start low.
Distance from the bottom, at dispenseA Z height where every droplet catches the well. Over cells: far enough that the tips do not scrape the cells, close enough that the droplet dips in the well.Start low. Over cells, higher.

There are no exact speeds or heights here. They differ with the instrument, the tips, the labware and the liquid, so those numbers are yours to set.

You do not always need an optimizer. Instead of fine-tuning with Bayesian optimization, use an expert-guided approach, and fine-tune to the level that is appropriate for the biological material.

The rulebook

The same parameters, in code.

These go into a rulebook. The agent says what the material is. The rulebook sets the speed, the heights and the blowout.

code/principles.py
"""Expert-guided fine-tuning, as a rulebook.

You do not always need an optimizer. Start with regular parameters and
maximum speed for everything, then fine-tune to the level that is
appropriate for the biological material.

These are the tunable parameters: the blowout volume, the speed and
the distance from the bottom. Use them as a starting principle for
what would work, then tune them on your own instrument.

There are no exact speeds or heights here. Speed is a share of the
top speed. The top speed and the heights differ with the instrument,
the tips, the labware and the liquid, so they are yours to set. A
real run needs them.

    python principles.py
"""
from dataclasses import dataclass
from typing import Optional

TOP_SPEED = None              # uL/s. Yours to set.
ASPIRATE_HEIGHT = None        # mm above the bottom. Yours to set.
DISPENSE_HEIGHT = None        # mm above the bottom. Yours to set.
CELLS_DISPENSE_HEIGHT = None  # mm above the bottom. Yours to set.
BLOWOUT_SHARE = 0.20          # blowout air, as a share of the volume
CELLS_SPEED_SHARE = 0.60      # 50 to 70 percent of the top speed

EVERY_TRANSFER = (
    "Start with regular parameters, maximum speed for everything.",
    "Sufficient blowout volume, to ensure proper liquid handling.",
    "Aspirate low enough that the liquid is not missed.",
)
CATCH = "Dispense at a Z height where every droplet catches the well."


@dataclass(frozen=True)
class Handling:
    aspirate_speed: float              # a share of the top speed
    aspirate_height: Optional[float]   # mm above the bottom
    dispense_speed: float              # a share of the top speed
    dispense_height: Optional[float]   # mm above the bottom
    why: tuple


def handling(material: str) -> Handling:
    """What the expert would set for this material."""
    if material == "regular":
        speed, height, why = 1.0, DISPENSE_HEIGHT, (CATCH,)
    elif material == "cells":
        speed = CELLS_SPEED_SHARE
        height = CELLS_DISPENSE_HEIGHT
        why = ("Gentler into cells: 50 to 70 percent of maximum speed.",
               "Far enough that the tips do not scrape the cells.",
               "Close enough that the droplet dips in the well.")
    elif material == "lysis":
        speed, height = 1.0, DISPENSE_HEIGHT
        why = ("A forceful dispense for lysis.", CATCH)
    else:
        raise ValueError(
            f"No principle for {material!r}. Ask the expert.")
    return Handling(1.0, ASPIRATE_HEIGHT, speed, height,
                    EVERY_TRANSFER + why)


def not_set(material: str = "") -> str:
    """Say which of your numbers this material still needs."""
    names = ["TOP_SPEED", "ASPIRATE_HEIGHT", "DISPENSE_HEIGHT",
             "CELLS_DISPENSE_HEIGHT"]
    if material == "cells":
        names.remove("DISPENSE_HEIGHT")
    elif material:
        names.remove("CELLS_DISPENSE_HEIGHT")
    missing = [name for name in names if globals()[name] is None]
    if not missing:
        return ""
    return ("Not set: " + ", ".join(missing) + ".\n"
            "They differ with the instrument, the tips, the labware and"
            " the\nliquid. Set yours in principles.py.")


def _args(share: float, height: Optional[float], vols):
    how = {"blow_out_air_volume": [BLOWOUT_SHARE * v for v in vols]}
    if height is not None:
        how["liquid_height"] = [height] * len(vols)
    if TOP_SPEED is not None:
        how["flow_rates"] = [share * TOP_SPEED] * len(vols)
    return how    # what is not set is left to PyLabRobot


def aspirate_args(h: Handling, vols):
    return _args(h.aspirate_speed, h.aspirate_height, vols)


def dispense_args(h: Handling, vols):
    return _args(h.dispense_speed, h.dispense_height, vols)


def speed(share: float) -> str:
    words = ("top speed" if share == 1
             else f"{share * 100:g} percent of top speed")
    if TOP_SPEED is None:
        return words
    return f"{words}, {share * TOP_SPEED:g} uL/s"


def height(mm: Optional[float]) -> str:
    if mm is None:
        return "height not set"
    return f"{mm:g} mm above the bottom"


def describe(material: str, volume: float) -> str:
    h = handling(material)
    lines = [
        f"{material}, {volume:g} uL",
        f"  aspirate at {speed(h.aspirate_speed)},"
        f" {height(h.aspirate_height)}",
        f"  dispense at {speed(h.dispense_speed)},"
        f" {height(h.dispense_height)}",
        f"  blowout  {BLOWOUT_SHARE * volume:g} uL",
    ]
    return "\n".join(lines + [f"  - {reason}" for reason in h.why])


if __name__ == "__main__":
    for material in ("regular", "cells", "lysis"):
        print(describe(material, 100))
        print()
    if not_set():
        print(not_set())
run it
python principles.py
what it prints
regular, 100 uL
  aspirate at top speed, height not set
  dispense at top speed, height not set
  blowout  20 uL
  - Start with regular parameters, maximum speed for everything.
  - Sufficient blowout volume, to ensure proper liquid handling.
  - Aspirate low enough that the liquid is not missed.
  - Dispense at a Z height where every droplet catches the well.

cells, 100 uL
  aspirate at top speed, height not set
  dispense at 60 percent of top speed, height not set
  blowout  20 uL
  - Start with regular parameters, maximum speed for everything.
  - Sufficient blowout volume, to ensure proper liquid handling.
  - Aspirate low enough that the liquid is not missed.
  - Gentler into cells: 50 to 70 percent of maximum speed.
  - Far enough that the tips do not scrape the cells.
  - Close enough that the droplet dips in the well.

lysis, 100 uL
  aspirate at top speed, height not set
  dispense at top speed, height not set
  blowout  20 uL
  - Start with regular parameters, maximum speed for everything.
  - Sufficient blowout volume, to ensure proper liquid handling.
  - Aspirate low enough that the liquid is not missed.
  - A forceful dispense for lysis.
  - Dispense at a Z height where every droplet catches the well.

Not set: TOP_SPEED, ASPIRATE_HEIGHT, DISPENSE_HEIGHT, CELLS_DISPENSE_HEIGHT.
They differ with the instrument, the tips, the labware and the
liquid. Set yours in principles.py.

The top speed and the heights are yours to set, and a real run does not start without them. Have your agent show you what your instrument and your labware allow.

Say to Claude Code

Read the PyLabRobot source you installed. What speeds does it allow for my instrument and my tips, and what does its own liquid class use? How deep are the wells of my labware? Show me where you read each number. I will pick the top speed and the heights.

To use the rulebook in your own harness, hand it to your agent.

Say to Claude Code

Read code/principles.py in the Agentic Lab Guide. Use it as the starting point for every transfer you write for my instrument. Before anything runs, show me the blowout, the speed and the distance from the bottom for each step, and tell me which numbers I still have to set.

Check it

Principles first, then the Rhodamine QC.

These principles first, then the Rhodamine QC. If it looks good, it is good for biological validation.

The QC makes the liquid something a plate reader can see, and gives you one number: the CV of the liquid transfer. To tune, change one parameter, run the plate, and read it again.

Where it goes next

Chapter 01 of the guide applies these on a Hamilton STAR. Layer 3 is the Rhodamine QC.